SK5 N-BUTANOL
SENSITIVITY KIT

EN13725 Odor Panelist N-Butanol Screening

The prepared mixture in the sensitivity kit is used to screen panelists with the Scentroid SM100 or any other olfactometer in accordance to standard EN13725

Most Utilized Industries

Screen Shot 2022-08-15 at 7.24_edited.png

Travel

The N-butanol sensitivity kit provides an excellent alternative to pressurized reference gas cylinders, especially for air travel

Screen Shot 2022-08-15 at 7.04_edited.png

Odour Expert Guidance

Need further assistance? Let our odour experts guide you each step of the way

Screen Shot 2022-08-15 at 7.03_edited.png

Multiple Concentrations

The N-butanol kit can be used to create any concentration from 4 to 40 ppm. Variations in concentration allows the lab to better test their equipment and the panelists

Screen Shot 2022-08-09 at 8_edited.png

Light and Portable Design

The reduced size and weight of the kit ensures that you can take on any odour adventure

Screen Shot 2022-08-12 at 4.25_edited.png

Cost-Effective Solution

The N-butanol kit provides an easy to use and cost effective solution to test panelist sensitivity training in accordance to all international standards, including EN13725

Screen Shot 2022-08-15 at 12.28_edited.png

Each kit contains: a micro-syringe, PTFE sample bag, stainless steel evaporation chamber, fitting, PTFE tube, hard carrying case

Related Products

Basics of Operation

  • Draw the required amount of liquid N-butanol according to the calibration chart

  • Inject the N-butanol into the evaporation chamber

  • Fill the evaporation chamber with odourless air from the SM100

  • Close the valve and wait for 15 minutes

  • Fill the 1L Tedlar bag

Overview

The Scentroid SK5 EN13725 Odor Panelist Sensitivity Kit is designed to provide accurate N-butanol mixtures of 5 to 50 ppm from a liquid N-butanol source. The N-butanol mixture can then be used to screen and odour panellist using the Scentroid SM100 or any other Scentroid olfactometer in accordance with standard EN13725. The mixture is accurate and repeatable thanks to the fixed volume evaporation chamber, micro-litre syringe, and pre-purged Tedlar bag.

Every odour regulation requires the screening of the odour panellist using diluted N-butanol samples. Until now, the only available source of a diluted N-butanol sample was purchased pre-mixed pressurized cylinders. The cylinders were cumbersome to transport, required a special permit and storage facility, and would only be accurate in dilution if shaken or rolled just before use. The Scentroid SK5 Sensitivity kit provides an easy and reliable solution for creating exact N-butanol samples that can be used for panellist screening. Liquid N-butanol is mixed with clean odourless air at a fixed ratio using a glass micro-syringe and a stainless steel evaporation chamber.

The N-butanol kit is easy to transport and store, does not have expiration issues, and allows variable concentration.

Each SK5 N-Butanol Kit Contains

  • Micro-syringe

  • PTFE sample bag

  • Stainless steel evaporation chamber

  • 1/4 fitting

  • PTFE tube

  • Hard carrying case

Calibration Chart